Photopeak efficiency response function of an underwater gamma ray NaI(TI) detector using MCNP X
2015
This work presents a study to calculate the response function of a 1.5”x 1” NaI(Tl) scintillation detector when it is used in the marine environment in the energy range from 20 keV to 662 keV. The method takes into account both the scattering of photons in the water and the detection mechanism of the detector. In addition, the calculation of the response function of the whole system is essential for suppressing the background of the measurement and for estimating the concentration of the involved radionuclides, especially given the greater probability of primary gamma photons undergoing multiple scattering events before they interact with the detector. The experimental photopeak efficiency measurements for point sources were compared with the simulated results under the same conditions of the experimental setup to validate the simulation of the detector. Monte Carlo simulations were performed using the MCNP-X code for the investigation of gamma-ray absorption in water in different brines. The energy resolution curve was used to improve the response of the mathematical simulation of the detector. The detector’s simulation was based on information obtained from the gammagraphy technique. Both dimensions and materials were used for the calculation with the MCNP-X code. The photopeak efficiency of a NaI(Tl) detector for different radionuclides in the aquatic environment with different salinities was calculated.