Experimental Investigation of the Accuracy of an Ultrawideband Time-Domain Microwave-Tomographic System

2011 
The measurement accuracy of an ultrawideband (UWB) time-domain microwave-tomographic system is investigated. In order to make an assessment of the random variation of the measurements, the measurement repeatability of the system is evaluated by comparison with an UWB frequency-domain system. A phantom is imaged with the time-domain microwave-tomographic system, and the reconstructed images are compared with those obtained by using the frequency-domain system. The results suggest that with the averaging tens of measurements, the time-domain system can achieve the same level of measurement repeatability as that of the frequency-domain system in the interesting frequency range of microwave tomography. The imaging results, however, indicate that the phantom reconstruction does not require such high measurement accuracy. The permittivity profile of the phantom reconstructed from the nonaveraging time-domain measurements is very similar with that obtained by means of the frequency-domain system.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    33
    References
    40
    Citations
    NaN
    KQI
    []