Instrument for precision long-term β-decay rate measurements

2015 
We describe an experimental setup for making precision measurements of relative β-decay rates of 22Na, 36Cl, 54Mn, 60Co, 90Sr, 133Ba, 137Cs, 152Eu, and 154Eu. The radioactive samples are mounted in two automated sample changers that sequentially position the samples with high spatial precision in front of sets of detectors. The set of detectors for one sample changer consists of four Geiger-Muller (GM) tubes and the other set of detectors consists of two NaI scintillators. The statistical uncertainty in the count rate is few times 0.01% per day for the GM detectors and about 0.01% per hour on the NaI detectors. The sample changers, detectors, and associated electronics are housed in a sealed chamber held at constant absolute pressure, humidity, and temperature to isolate the experiment from environmental variations. The apparatus is designed to accumulate statistics over many years in a regulated environment to test recent claims of small annual variations in the decay rates. We demonstrate that absent th...