Prediction of electron beam output factors

1982 
A method to predict square and rectangular field output factors from the measurement of selected fields of electron beams on the Therac 20 Saturne has been developed. A two parameter fit of the square field output factor data, based on the functional dependence as predicted by a pencil beam calculational model, has proven clinically acceptable. The pencil beam distributions are given by the Fermi–Eyges theory of multiple Coulomb scattering. For a rectangular field, the output factor can be calculated from the square root of the product of the two square field output factors with sides equal to those of the rectangular field. If however, there is a significant asymmetry between the X and Y collimator systems, then rectangular field output factors should be predicted from the product of the X and Y one‐dimensional output factors. One‐dimensional output factors are defined as output factors of rectangular fields where one side remains constant and equal to the side of the square reference field. Measured data indicate either of the two methods of determining rectangular field output factors to be clinically acceptable for the Therac 20, the use of one‐dimensional output factors demonstrating greater accuracy. Data show agreement to within approximately 1.5% at electron energies of 6, 9, 13, and 17 MeV.