Validity of the VmaxST portable metabolic measurement system

2008 
Abstract The aim of this study was to assess and compare the validity of the portable VmaxST telemetry metabolic measurement device with that of a standard measurement system (Vmax29). Thirty asymptomatic, moderately active males provided written, informed consent and completed two maximal graded treadmill exercise tests (Bruce) using the VmaxST and the Vmax29 metabolic measurement systems. Tests were performed in random order on separate days to obtain peak values for time to exhaustion, heart rate, systolic and diastolic blood pressure, oxygen consumption ([Vdot]O2), carbon dioxide production ([Vdot]CO2), ventilation ([Vdot] E), and respiratory exchange ratio (RER). Multivariate analysis of variance revealed no significant main effect (P = 0.88) between the two systems across any variable, suggesting similar measurement capabilities between the two systems. Linear regression analyses revealed moderate to high coefficients of determination for [Vdot]O2 (r 2 = 0.99), [Vdot]CO2 (r 2 = 0.99), [Vdot] E (r 2 ...