Note: Pneumatically modulated liquid delivery with feedback control

2012 
We present the design and characterization of a pneumatically driven liquid delivery system using an embedded microcontroller with feedback control capable of maintaining a stable, constant flow rate over several hours of operation. Flow rates with relative standard deviations less than 1% were achieved and compared to a typical laboratory syringe pump.