Do Serum Creatinine Levels Show Clinically Significant Fluctuations on Serial Determinations on the Siemens Advia 1800 Analyzer
2017
Background
The goal of this work was to determine whether there are clinically significant fluctuations in the level of serum creatinine on serial determinations, especially in the borderline range (1.1–1.3 mg/dl), after specimen storage.
Methods
Sixty-one serum samples were analyzed. They were divided into three categories based on the initial serum creatinine measurement: low (≤1.0 mg/dl), borderline (1.1–1.3 mg/dl), and high (≥1.4 mg/dl). The specimens were stored at 4°C and run on the Siemens Advia 1800 chemistry analyzer on days 1, 3, and 11.
Results
Statistical comparisons of the three groups were made using the unpaired t-test, yielding a two-tailed P-value for each group comparison. The P-values ranged from 0.0829 to 0.3892, indicating no statistically significant difference between the standard deviations of each group.
Conclusions
Mild-to-moderate fluctuations in precision occur in successive serum creatinine determinations. The overwhelming majority of these fluctuations should not affect clinical decision making.