Accurate method to quantify binding in supramolecular chemistry

2013 
An approach for accurate and comparable measurement of host–guest binding affinities is introduced whereby differences in binding strength (ΔlogKass values) are measured between two host molecules toward a particular guest under identical solvent conditions. Measuring differences instead of absolute values enables obtaining highly accurate results, because many of the uncertainty sources (the solvation/association state of the guest in solution, deviations in solvent composition, etc.) cancel out. As a proof of concept, this method was applied to the measurement of the binding strength of 28 synthetic anion receptors toward acetate in acetonitrile containing 0.5% water. The receptors included differently substituted indolocarbazoles, ureas, thioureas, and some others. Possible deprotonation of more acidic receptors of each compound class by acetate was checked by measuring their acidities (ΔpKa values) relative to acetic acid in the same solvent. A self-consistent (consistency standard deviation 0.04 log ...