Background An appropriate size of device for patent ductus arteriosus (PDA) could be chosen by a stretched PDA size. We propose prediction of stretched size from intact PDA size. Methods A total of 361 patients was enrolled. Intact size was measured on angiography before closure, and stretched size was the narrowest diameter immediately after device implantation. Results The median patient age was 2.8 years. Intact diameter and stretched diameter were 3.6 ± 1.4 mm and 5.1 ± 1.3 mm. The difference and difference ratio were 1.5 ± 0.8 mm and 51.3 ± 38.1% and had negative linear correlations with age (P Conclusions Stretched PDA diameter can be predicted from intact diameter and might be helpful for safe PDA closure especially in small infants.