Mechanical characterization and comparison of energy storage and return prostheses

2017 
Abstract The suitability of finite element analysis (FEA) for standardizing the mechanical characterization of energy storage and return (ESAR) prostheses was investigated. A methodology consisting of both experimental and numerical analysis was proposed and trialed for the Vari-flex ® Modular TM , Flex-foot Cheetah and Cheetah Xtreme by Ossur ® and a 1E90 Sprinter by Ottobock ® . Gait analysis was conducted to determine suitable orientation angles for non-destructive testing (NDT) of the ESAR prostheses followed by a quasi-static inverse FEA procedure within COMSOL Multiphysics ® , where the NDT conditions were replicated to determine the homogenized material properties of the prostheses. The prostheses' loading response under bodyweight for an 80kg person was then simulated, using both Eigenfrequency and time-dependent analysis. The apparent stiffness under bodyweight was determined to be 94.7, 48.6, 57.4 and 65.0Nmm −1 for the Vari-flex ® Modular TM , Flex-foot Cheetah, Cheetah Xtreme and 1E90 Sprinter, respectively. Both the energy stored and returned by the prostheses varied negatively with stiffness, yet the overall efficiency of the prostheses were similar, at 52.7, 52.0, 51.7 and 52.4% for the abovementioned prostheses. The proposed methodology allows the standardized assessment and comparison of ESAR prostheses without the confounding influences of subject-specific gait characteristics.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    24
    References
    11
    Citations
    NaN
    KQI
    []