The use of a ceramic talar body prosthesis in patients with aseptic necrosis of the talus
2012
The purpose of this study was to evaluate the
clinical results of a newly designed prosthesis to replace the body
of the talus in patients with aseptic necrosis. Between 1999 and
2006, 22 tali in 22 patients were replaced with a ceramic prosthesis.
A total of eight patients were treated with the first-generation
prosthesis, incorporating a peg to fix into the retained neck and
head of the talus, and the remaining 14 were treated with the second-generation prosthesis,
which does not have the peg. The clinical results were assessed
by the American Orthopaedic Foot and Ankle Society ankle/hindfoot
scale. The mean follow-up was 98 months (18 to 174). The clinical results
of the first-generation prostheses were excellent in three patients,
good in one, fair in three and poor in one. There were, however,
radiological signs of loosening, prompting a change in design. The
clinical results of the second-generation prostheses were excellent
in three patients, good in five, fair in four and poor in two, with
more favourable radiological appearances. Revision was required
using a total talar implant in four patients, two in each group. Although the second-generation prosthesis produced better results,
we cannot recommend the use of a talar body prosthesis. We now recommend
the use of a total talar implant in these patients.