Cryoablation with an 8-mm-Tip Catheter for Right-Sided Accessory Pathways in Children.

2016 
Background Cryoablation is increasingly utilized in children because of its safety profile. Recently, larger catheter tips have been more widely used to improve long-term success rates. The aim of this study was to assess the safety and efficacy of 8-mm-tip catheters for cryoablation of right-sided accessory pathways (APs) in children. Methods Electrophysiological procedures were performed using the EnSite™ system (St. Jude Medical Inc., St. Paul, MN, USA). Results Between July 2010 and July 2014, 54 patients (mean age: 13.1 ± 3.7 years) underwent cryoablation using an 8-mm-tip catheter. In 18 of 54 (33%) patients where an 8-mm-tip catheter was the first-choice catheter, the success rate was 18 of 18 (100%). There was a history of previous failed attempts or recurrence with radiofrequency ablation and/or 6-mm-tip cryoablation in 36 of 54 (67%) patients. The success rate in these patients was 24 of 36 (67%). No fluoroscopy was used in 34 of 54 procedures. The recurrence rate was six of 42 (14%) during a mean follow-up period of 32 ± 15 months. In one patient, transient atrioventricular block occurred. Conclusions Cryoablation with an 8-mm-tip catheter for right-sided APs in children who weigh over 40 kg appears to be safe and acutely effective in cases where conventional ablation methods fail and also as a first choice for ablation procedure. However, the recurrence rate still seems to be high.