Monitoring optical properties of aerosols with cavity ring-down spectroscopy

2011 
Abstract This article describes the design and performance of a cavity ring-down spectroscopic (CRDS) instrument for measuring extinction coefficients of laboratory and ambient atmospheric aerosols. Through averaging 1000 individual waveforms, a minimum detectable aerosol extinction coefficient of 6.1×10 −7  m −1 is achieved. Tests with polystyrene spheres (PSS), we suggested this CRDS system could measure the extinction coefficient of aerosol with uncertainty x analyzer, CRDS was used to monitor the optical properties of ambient aerosols in the heavy pollution episode. The uncertainty for using the CRDS and TSI nephelometer to measure single scattering albedo (SSA) in an ambient measurement is estimated to be