Evaluation of three hollow-fiber membrane oxygenators without integrated arterial filters for neonatal cardiopulmonary bypass

2012 
The cardiopulmonary bypass (CPB) procedure has been shown to be a possible cause of postoperative neurological morbidity for various reasons, including: large amounts of gaseous microemboli (GME) reaching the patient and hypoperfusion of the patient due to “stolen” blood flow. This study used a simulated CPB circuit identical to that in a clinical setting to examine three different hollow-fiber membrane oxygenators without intergrated arterial filters - the Capiox RX05, the Quadrox-i neonatal, and the KIDS D100 - to determine their ability to reduce the number of GME delivered to the neonatal patient and their hemodynamic properties in response to varying flow rates, normothermic vs hypothermic conditions, and open vs closed purge line. The circuit was primed with Ringer’s Lactate and then human blood with a hematocrit of 30%. Injections of 5cc bolusses of air were injected into the venous line proximal to the venous reservoir over a thirty-second interval. Six injections were done for each oxygenator at ...