Large microcalorimeter arrays for high-resolution X- and gamma-rayspectroscopy

2011 
Abstract Microcalorimeter detectors provide unprecedented energy resolution for the measurement of X-rays and soft gamma-rays. Energy resolution in the 100 keV region can be up to an order of magnitude better than planar high-purity germanium (HPGe) detectors. The technology is well-suited to analysis of materials with complex spectra presenting closely spaced photopeaks. One application area is the measurement and assay of nuclear materials for safeguards and fuel cycle applications. In this paper, we discuss the operation and performance of a 256-pixel array, and present results of a head-to-head comparison of isotopic determination measurements with high-purity germanium using a plutonium standard. We show that the uncertainty of a single measurement is smaller for the microcalorimeter data compared to the HPGe data when photopeak areas are equal. We identify several key areas where analysis codes can be optimized that will likely lead to improvement in the microcalorimeter performance.