Regional hyperthermia for deep-seated malignancies using the BSD annular array
1988
Forty-four patients were treated using the BSD-1000 Annular Phased Array between April 1983 and December 1986. There were 32 pelvic, nine abdominal, two extremity, and one thoracic sites treated. Mean tumour volume was 646 cc. Thirty-nine patients had concurrent radiation therapy, receiving a mean dose of 38 Gy. Mean average temperature was 41.0 ± 1.4°C. Most patients experienced local or systemic toxicity, requiring temporary treatment interruption in 33 patients, and termination of treatment in eight. Chronic complications were seen in four, but these were in patients receiving high total radiation doses as well. There were six complete and five partial responses.Among the 32 patients with pelvic tumours, mean tumour volume was 317 cc, mean radiation dose was 42 Gy, and mean average temperature was 41.3 ± 1.2°C. There were five complete and four partial responses.Achieving tumour temperatures ± 42°C with the annular array is difficult, due to both systemic and local toxicity. To improve clinical hyperth...