Comparing the accuracy of transcutaneous sensor and 90-DAY implantable glucose sensor

2020 
Abstract Background and Aims Continuous glucose monitoring improves glycemic control in diabetes. This study compared the accuracy of the Dexcom G5 Mobile (Dexcom, San Diego, CA) transcutaneous sensor (DG5) and the first version of Eversense (Senseonics,Inc., Germantown, MD) implantable sensor (EVS). Methods and results Subjects with type 1 diabetes (T1D) and using EVS wore simultaneously DG5 for seven days. At day 3, patients were admitted to a clinical research center (CRC) to receive breakfast with delayed and increased insulin bolus to induce glucose excursions. At CRC, venous glucose was monitored every 15 min (or 5 min during hypoglycemia) for 6 hours by YSI 2300 STAT PLUSTM glucose and lactate analyzer. At home patients were requested to perform 4 fingerstick glucose measurements per day. 11 patients (9 males, age 47.4±11.3 years, M±SD) were enrolled. During home-stay the median [25th-75th percentile] absolute relative difference (ARD) over all CGM-fingerstick matched-pairs was 11.64% [5.38-20.65]% for the DG5 and 10.75% [5.15-19.74]% for the EVS (p-value=0.58). At CRC, considering all the CGM-YSI matched-pairs, the DG5 showed overall smaller median ARD than EVS, 7.91% [4.14-14.30]% vs 11.4% [5.04- 18.54]% (p-value Conclusions DG5 was more accurate than EVS at CRC, especially when glucose decreased. No differences were found at home.