Liver MRI at 3.0 tesla: comparison of image quality and lesion detectability between single-source conventional and dual-source parallel radiofrequency transmissions.

2012 
OBJECTIVE: To prospectively and intraindividually compare liver magnetic resonance imaging (MRI) using single-source and dual-source parallel radiofrequency (RF) transmissions at 3.0-T for image quality, lesion detectability, and lesion contrast. METHODS: Ninety-nine patients with 139 liver lesions underwent liver MRI at 3.0-T. Two radiologists performed a consensus review of T2-weighted images (T2WI), heavily T2WI (HT2WI), gadoxetic acid-enhanced hepatobiliary images, and diffusion-weighted imaging using single-source and dual-source RF transmissions with regard to image quality and lesion detectability. Contrast ratios between liver lesions and liver parenchyma were also calculated. RESULTS: Image quality was better with dual-source than with single-source at T2WI and HT2WI (P 0.05). CONCLUSION: Dual-source RF transmission provides a better image quality with T2WI and HT2WI than with single-source. However, 2 techniques showed similar lesion detectability.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    25
    References
    8
    Citations
    NaN
    KQI
    []