Magnetic colorimetric immunoassay for human interleukin-6 based on the oxidase activity of ceria spheres.
2016
Abstract A novel magnetic colorimetric immunoassay strategy was designed for sensitive detection of human interleukin-6 (IL-6) using ceria spheres as labels. Ceria spheres showed excellent oxidase activity, which can directly catalyze the oxidation of substrate o-phenylenediamine (OPD) to a stable yellow product, 2,3-diaminophenazine (oxOPD). The absorbance of oxOPD was recorded to reflect the level of IL-6. The relatively mild conditions made the immunoassay strategy more robust, reliable, and easy. A linear relationship between absorbance intensity and the logarithm of IL-6 concentrations was obtained in the range of 0.0001–10 ng mL −1 with a detection limit of 0.04 pg mL −1 (S/N = 3). The colorimetric immunoassay exhibited high sensitivity and specificity for the detection of IL-6. This immunoassay has been successfully applied in the detection of IL-6 in serum samples and can be readily extended toward the on-site monitoring of cancer biomarkers in serum samples.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
42
References
20
Citations
NaN
KQI