Insufficiency in a New Temporal-Artery Thermometer for Adult and Pediatric Patients

2002 
SensorTouch™ is a new noninvasive temperature monitor and consists of an infrared scanner that detects the highest temperature on the skin of the forehead, presumably over the temporal artery. The device estimates core temperature (Tcore). We tested the hypothesis that the SensorTouch™ is sufficiently precise and accurate for routine clinical use. We studied adults (n 15) and children (n 16) who developed mild fever, a core temperature of at least 37.8°C, after cardiopulmonary bypass. Temperature was recorded at 15-min intervals throughout recovery with the SensorTouch™ thermometer and from the pulmonary artery (adults) or bladder (children). Pulmonary artery (Tcore) and SensorTouch™ (Tst) temperatures correlated poorly in adults: Tcore 0.7 Tst 13, r 2 0.3. Infrared and pulmonary artery temperatures differed by 1.3 0.6°C; 89% of the adult temperatures thus differed by more than 0.5°C. Bladder and infrared temperatures correlated somewhat better in pediatric patients: Tcore 0.9 Tst 12, r 2 0.6. Infrared and bladder temperatures in children differed by only 0.3°C, but the sd of the difference was 0.5°C. Thus, 31% of the values in the infants and children differed by more than 0.5°C. (Anesth Analg 2002;95:67–71)