Diagnostic and analytical study of post ablation ionization of neutral atoms of major and minor constituents from a low alloy steel

1996 
Post ablation ionization (PAI) of neutral atoms from a low alloy steel has been investigated using non-resonant laser ionization in a time-of-flight mass spectrometer. By varying the delay between the ablation and ionization lasers, the velocity distributions of the Ti, V, Cr, Mn and Fe atoms have been determined simultaneously. These distributions have been recorded as a function of ablation laser fluence. The half-range Maxwell-Boltzmann velocity distribution has been used to fit the data and different characteristic temperatures have been determined for the various elements in the sample. The quantitative capability of this method for bulk and surface analysis has been evaluated by calculating the relative sensitivity factors (RSFs) for the various constituent elements. The RSFs for all of the elements are seen to be highly dependent on the delay between the ablating and ionizing lasers. This dependence was reduced by integrating the temporal dependent ion yield, leading to a significant improvement in the calculated RSF values. It was also found that the RSFs were not highly dependent on the power density of the ablation laser beam.