An Evaluation of Portable Wet Bulb Globe Temperature Monitor Accuracy

2017 
Context:  Wet bulb globe temperature (WBGT) is the gold standard for assessing environmental heat stress during physical activity. Many manufacturers of commercially available instruments fail to report WBGT accuracy. Objective:  To determine the accuracy of several commercially available WBGT monitors compared with a standardized reference device. Design:  Observational study. Setting:  Field test. Patients or Other Participants:  Six commercially available WBGT devices. Main Outcome Measure(s):  Data were recorded for 3 sessions (1 in the morning and 2 in the afternoon) at 2-minute intervals for at least 2 hours. Mean absolute error (MAE), root mean square error (RMSE), mean bias error (MBE), and the Pearson correlation coefficient (r) were calculated to determine instrument performance compared with the reference unit. Results:  The QUESTemp° 34 (MAE = 0.24°C, RMSE = 0.44°C, MBE = –0.64%) and Extech HT30 Heat Stress Wet Bulb Globe Temperature Meter (Extech; MAE = 0.61°C, RMSE = 0.79°C, MBE = 0.44%) dem...