Cranioplasty with Customized Titanium and PEEK Implants in a Mechanical Stress Model
2012
Abstract Large skull defects as a result of craniectomies due to cerebral insults, trauma, or tumors create functional and aesthetic disturbances for the patient. Cranioplasty with implants in these cases are an alternative to autogenous bone transplantation. In our clinic, customized titanium or optima poly-ether-ether ketone (PEEK) implants are used to reconstruct craniectomy defects. To compare the two materials we investigated the structural changes of the implants fixed to a sintered polyamide skull model under mechanical stress in four simplified models. In a standard testing machine, the models were subjected to a load under a quasi-static loading rate of 1.925 mm/min. Fractures of the PEEK implants occurred at a force of 24.2 and 24.5 kN with a displacement of 8.4 and 8 mm. The titanium implants showed no deformation, but extensive damage was seen in the polyamide skull models. The highest pressures achieved were 45.8 and 50.9 kN. In a simplified model with quasi-static loading, both implants with...