Final report of the SIM 60Co absorbed-dose-to-water comparison SIM.RI(I)-K4

2008 
Transfer chambers were used to compare the standards for 60Co absorbed dose to water maintained by seven laboratories. Six of the laboratories were members of the Sistema Interamericano de Metrologia (SIM) regional metrology organization while the seventh was the International Atomic Energy Agency (IAEA) laboratory in Vienna. The National Research Council (NRC) acted as the pilot laboratory for the comparison. Because of the participation of laboratories holding primary standards, the comparison results could be linked to the key comparison reference value maintained by the Bureau International des Poids et Mesures (BIPM). The results for all laboratories were within the expanded uncertainty (two standard deviations) of the reference value. The estimated relative standard uncertainty on the comparison between any pair of laboratories ranged from 0.6% to 1.4%. The largest discrepancy between any two laboratories was 1.3%. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).