Accuracy of alveolar bone height and thickness measurements in cone beam computed tomography: a systematic review and meta-analysis

2019 
ABSTRACT Objective The aim of this study was to systematically review and assess the accuracy of cone beam computed tomography (CBCT) in the measurements of alveolar bone height and thickness. Study Design MEDLINE, Embase, the Cochrane Library, China National Knowledge Infrastructure, and the grey literature were searched to identify all relevant articles published prior to July 2018. The Quality Assessment of Measurement Accuracy Studies (QUAMAS) tool was used to assess the quality of the included studies. Meta-analysis was performed to analyze the mean differences of the alveolar bone height and thickness measurements between CBCT and gold standard references (direct measurement on human skulls or live patients). Results In total, 28 studies were included in the qualitative synthesis and 18 were included in the quantitative synthesis. The meta-analysis results showed that the mean differences between CBCT measurements and the gold standard references for alveolar bone height (mean difference = 0.03mm; 95% CI: -0.03 to 0.08, p=0.382) and alveolar bone thickness (mean difference = 0.11mm; 95% CI: -0.02 to 0.24, p=0.088) were not statistically significant. Conclusions Current evidence suggests that there is no significant difference between CBCT and the gold standard references for the measurement of alveolar bone height and thickness.