Characterization of Polycapillary Optics in a TES Microcalorimeter EDS System Installed on an SEM
2016
Energy-dispersive spectroscopic measurements are performed using a superconducting transition-edge sensor (TES) microcalorimeter mounted on a scanning electron microscope (SEM) for advanced research at Kyushu University. Because the sensitive area of the TES microcalorimeter is about \(0.02~\mathrm{mm}^{2}\), polycapillary optics is used to collect the X-rays emitted by the SEM specimen on the TES microcalorimeter. The X-ray transmission efficiency of the polycapillary optics is obtained by analyzing the X-ray energy spectra measured by the TES microcalorimeter. The obtained transmission efficiency of the polycapillary optics is reproduced by the calculated results of the simulation.