Operation of sealed microstrip gas chambers at the ILL

2001 
Abstract Microstrip Gas Counters (MSGCs) were introduced at the ILL as a response to the problem of fabricating the large area neutron detector of the D20 neutron powder diffractometer. This banana-like detector consists of 48 MSGCs, each comprising 32 counting cells. It was in operation during 18 months before being stopped due to the progressive deterioration of the anode strips. In order to increase its lifetime, significant modifications were introduced in the recently assembled new version. Another instrument, D4C, was recently equipped with a modular detector made of nine MSGCs, each of them in an individual gas vessel. Besides the unidimensional individual readout MSGC of D20 and D4C, the ILL has developed bidimensional MSGCs with a charge division readout. All these detectors employ sealed vessels containing a gas mixture at a pressure which can be as high as 15 bar, necessitating very clean conditions. This paper describes the experience acquired at the ILL in the fabrication and operation of these detectors.