Clinical evaluation of the air-handling properties of contemporary oxygenators with integrated arterial filter:

2017 
Gaseous microemboli (GME) may originate from the extracorporeal circuit and enter the arterial circulation of the patient. GME are thought to contribute to cerebral deficit and to adverse outcome after cardiac surgery. The arterial filter is a specially designed component for removing both gaseous and solid microemboli. Integration of an arterial filter with an oxygenator is a contemporary concept, reducing both prime volume and foreign surface area. This study aims to determine the air-handling properties of four contemporary oxygenator devices with an integrated arterial filter. Two oxygenator devices, the Capiox FX25 and the Fusion, showed significant increased volume of GME reduction rates (95.03 ± 3.13% and 95.74 ± 2.69%, respectively) compared with both the Quadrox-IF (85.23 ± 5.84%) and the Inspire 6F M (84.41 ± 12.93%). Notably, both the Quadrox-IF and the Inspire 6F M as well as the Capiox FX 25 and the Fusion showed very similar characteristics in volume and number reduction rates and in detaile...