Deficiencies of automatic endoscopic reprocessors: a method to achieve high-grade disinfection of endoscopes

1996 
Abstract Background: We show that disinfection using the automatic endoscopic reprocessor is not complete and propose a method for high-grade disinfection of endoscopes. Methods: We used an automatic endoscopic reprocessor, Pyser System 83, and 2% glutaraldehyde. After each endoscopic procedure, the endoscopes were divided into three groups. Endoscopes in group A were washed only by the reprocessor. Group B endoscopes were washed by the reprocessor after the connectors were soaked in glutaraldehyde for 5 minutes. The channels, valves, connecting sections of group C endoscopes, and the connectors of the machine were sprayed with glutaraldehyde before machine-washing. Swabs were taken from all 123 parts of each endoscope and machine for microbiologic cultrue. Results: Six endoscopes were positive, cumulatively, for bacterial contamination in group A. Among group B endoscopes, one remained contaminated. no endoscope was positive in group C. The difference between group A and C was stitistically significant ( p Conclusions: Machine wasing by automatic endoscopic reprocessors may not achieve complete disinfection. Additional procedures are necessary. High-grade disinfection of the connectors is critical. Disinfection of the interface between the connectors is important. (Gastrointest Endosc 1996;44:583-6.