Implementation to spanish protocol of quality control of accelerators to daily control of electron beams; Implementacion del protocolo espanol de control de calidad de aceleradores al control diario de haces de electrones

2011 
A revised procedure for daily control of the electron beams to make measurements more meaningful physically, having a better reproducibility and more in line with the recommendations of the Spanish Protocol for Quality Control in Electron Linear Accelerators Clinical Use. The daily quality control beams of high energy electrons that had been done so far was the finding that the record of a series of measures (symmetry, uniformity, stability, energy, beam central dose) were within tolerance values established. The amendment is to check the beam quality by directly measuring changes in absorption depth at which the dose is reduced to half its maximum value, R50.