Fast and sensitive flow-injection mass spectrometry metabolomics by analyzing sample-specific ion distributions.

2020 
Mass spectrometry based metabolomics is a widely used approach in biomedical research. However, current methods coupling mass spectrometry with chromatography are time-consuming and not suitable for high-throughput analysis of thousands of samples. An alternative approach is flow-injection mass spectrometry (FI-MS) in which samples are directly injected to the ionization source. Here, we show that the sensitivity of Orbitrap FI-MS metabolomics methods is limited by ion competition effect. We describe an approach for overcoming this effect by analyzing the distribution of ion m/z values and computationally determining a series of optimal scan ranges. This enables reproducible detection of ~9,000 and ~10,000 m/z features in metabolomics and lipidomics analysis of serum samples, respectively, with a sample scan time of ~15 s and duty time of ~30 s; a ~50% increase versus current spectral-stitching FI-MS. This approach facilitates high-throughput metabolomics for a variety of applications, including biomarker discovery and functional genomics screens. Flow-injection mass spectrometry (FI-MS) enables high-throughput metabolomic profiling, but ion overload typically limits its sensitivity. Here, the authors show rapid and highly sensitive FI-MS overcoming an overload of the Orbitrap by analyzing sample-specific ion distributions.