Stress Thallium-201/Rest Technetium-99m Sequential Dual Isotope High-Speed Myocardial Perfusion Imaging
2009
Objectives Our purpose was to describe a novel, rapid stress thallium-201 (Tl-201)/rest technetium-99m (Tc-99m) agent myocardial perfusion imaging (MPI) protocol (Tl/Tc) with a high-speed MPI scanner and to compare this protocol with a standard rest/stress Tc-99m agent protocol (Tc/Tc) with respect to image quality and radiation dosimetry. Background Recent advances in gamma camera technology have provided opportunity for improved SPECT MPI protocols. A rapid Tl/Tc protocol that could improve image information while maintaining a low radiation burden for the patient would be desirable. Methods We compared high-speed SPECT MPI studies in 374 consecutive patients undergoing exercise or pharmacologic Tl/Tc protocol to those of 262 patients undergoing rest/stress Tc/Tc protocol. Results Tl/Tc imaging was accomplished in Conclusions A rapid stress Tl-201/rest Tc-99m protocol for use with high-speed SPECT MPI has image quality and radiation dosimetry similar to those observed with a conventional rest/stress Tc-99m protocol. The Tl/Tc protocol offers promise as an efficient and relatively low radiation dose method, in which the superior qualities of Tl-201 for stress imaging and of the Tc-99m agents for rest imaging can be preserved. The findings also suggest that with rapid imaging rest MPI immediately after Tc-99m agent injection may be superior to standard delayed image initiation.