MCP-BASED PHOTON DETECTOR WITH EXTENDED WAVELENGTH RANGE FOR FLASH

2007 
Experimental experience gained at the extreme ultraviolet SASE FEL FLASH (DESY, Hamburg) has shown that successful operation of the facility strongly depends on the quality of the radiation detectors. Here key issues are: wide wavelength range (6 to 100 nm for FLASH), wide dynamic range (from the level of spontaneous emission to the saturation level), and high relative accuracy of measurements which is crucial for detection of a signature of amplification and characterization of statistical properties of the radiation. In this report we describe MCP-based radiation detector for FLASH which meets these requirements. Key element of the detector is wide dynamic range micro-channel plate (MCP) which detects scattered radiation from a target. With four different targets and MCPs in combination with optical attenuators present detector covers operating wavelength range from 6 to 100 nm, and dynamic range of the radiation intensities, from the level of spontaneous emission up to the saturation level of SASE FEL.