Use of autobilevel ventilation in patients with obstructive sleep apnea: An observational study

2018 
: Continuous positive airway pressure (CPAP) is the first-choice treatment for obstructive sleep-disordered breathing. Automatic bilevel ventilation can be used to treat obstructive sleep-disordered breathing when CPAP is ineffective, but clinical experience is still limited. To assess the outcome of titration with CPAP and automatic bilevel ventilation, the charts of 356 outpatients (obstructive sleep apnea, n = 242; chronic obstructive pulmonary disease + obstructive sleep apnea overlap, n = 80; obesity hypoventilation syndrome [OHS], n = 34; 103 females) treated for obstructive sleep-disordered breathing from January 2014 to April 2017 were reviewed. Positive airway pressure titration was considered successful in the case of sleep-disordered breathing resolution (apnea-hypopnea index 40 kg/m2 (odds ratio 6.16, confidence interval 1.50-25.17, p = 0.011) and CT90% >42% (odds ratio 5.87, confidence interval 1.39-24.83, p = 0.016). During follow-up, automatic bilevel ventilation treatment failed in seven patients (10%), and compliance was similar in CPAP (4.5 ± 2.2 hr) and automatic bilevel ventilation (5.2 ± 2.3 hr, p = 0.09) groups. Automatic bilevel ventilation was useful to treat sleep-disordered breathing, but failed in patients with severe OHS.