Compact and accurate concept of laser wavemeters based on ellipsometry.

2011 
Common laser wavemeters are based on a scanning Michelson interferometer. Displacements of the moving mirror as long as tens of centimeters are needed to reach a relative accuracy of 1 × 10−6 (1σ) on the unknown laser wavelengths. Such a long displacement range makes the system very sensitive to mechanical vibrations and to misalignments of the laser beams. The purpose of this paper is to demonstrate a new concept of laser wavemeter based on the measurements of the ellipsometric parameters ψ and Δ of the laser beams. Experimental results show that a 10−6 (1σ) accuracy level could be reach with a displacement range of only 4 μm. Implementations of the device are described. Comparisons between our polarimetric wavemeter and a calibrated wavemeter are presented for two lasers, an extended cavity laser diode at 656 nm and a 532 nm green line Nd:YAG laser.