Strontium Iodide Instrument Development for Gamma Spectroscopy and Radioisotope Identification

2014 
Development of the Europium-doped Strontium Iodide scintillator, SrI2(Eu2+), has progressed significantly in recent years. SrI2(Eu2+) has excellent material properties for gamma ray spectroscopy: high light yield ( 2+ degrades the energy resolution as measured by analog electronics, but we mitigate this effect through on-the-fly correction of the scintillation pulses by digital readout electronics. Using this digital correction technique we have demonstrated energy resolution of 2.9% FWHM at 662 keV for a 4 in3 SrI 2 (Eu) crystal, over 2.6 inches long. Based on this digital readout technology, we have developed a detector prototype with greatly improved radioisotope identification capability compared to Sodium Iodide, NaI(Tl). The higher resolution of SrI 2 (Eu) yields a factor of 2 to 5 improvement in radioisotope identification (RIID) error rate compared to NaI(Tl).
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    13
    References
    8
    Citations
    NaN
    KQI
    []