Results of the gamma-neutron mapper performance test on 55-gallon drums at the RWMC

1995 
The results from a performance test of a {gamma}- and neutron-radiation measurement instrument used to provide two-dimensional radiation field maps are reported. The performance test was conducted at the Transuranic Storage Area of the Radioactive Waste Management Complex (RWMC) where interim storage is provided for 55-gal. drums of TRU waste from the Department of Energy`s Rocky Flats Plant. The performance test consisted of scanning drums stacked five high and five wide to identify high radiation areas and possible discrepancies with the waste manifest. Scans were taken at standoff distances of 15 cm, 30 cm, 45 cm and 90 cm. Data were acquired at scan speeds of 7.5 cm/s and 15 cm/s. The results of these scans are presented as one, two and three dimensional contour plots of the radiation fields. A comparison of these results with manifests of these drums are compared and discussed. While the T-radiation fields as measured by the Health Physicist and by the radiation maps are in general in agreement, the TRU content as given in the manifest did not often correlate with the neutron map.