The Development of Phasemeter for Taiji Space Gravitational Wave Detection

2018 
Taiji space gravitational wave detection utilizes the laser interferometer to convert the tiny distance change into the phase fluctuation of the beat note. As to realize the sensitivity of 1 pm/√ Hz, the phasemeter needs to calculate the phase with the precision of 2πμ rad/√ Hz in the frequency range of 0.1 mHz and 1 Hz. In this paper, we report recent progress of the phasemeter for Taiji. Noises which possibly affect the measurement sensitivity are tested and discussed, especially the sampling noise and the frequency jitter. Finally, the accuracy of the phasemeter is calibrated. The result shows that the sensitivity has reached the requirement of Taiji in the frequencies between 0.01 Hz and 1 Hz, 0.1 mHz–1 mHz. Noises in the range of 1 mHz and 0.01 Hz, which have not yet depressed well, are dominated by the clocking jitter and the thermal fluctuation.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    35
    References
    14
    Citations
    NaN