Prospective Comparison of 70-kVp Single-Energy CT versus Dual-Energy CT: Which is More Suitable for CT Angiography with Low Contrast Media Dosage?

2019 
Rationale and Objectives To compare the objective and subjective image qualities between single-energy computed tomography (CT) at 70 kVp and virtual monoenergetic imaging (VMI) of dual-source dual-energy CT for CT angiography with 180 mgI/kg. Materials and Methods Total 63 patients scanned with 180 mgI/kg were randomly divided into two groups: Group A (32 patients) underwent CT angiography at 70-kVp, and Group B (31 patients) underwent dual-energy CT. VMI sets were generated at 10-keV increments between 40 and 100 keV. We calculated aortic attenuation, contrast-to-noise-ratio (CNR), signal-to-noise-ratio, figure of merit of CNR, and effective dose for each protocol. Three radiologists scored overall image quality and various arteries’ visibility using a four-point scale. Quantitative and qualitative comparisons between 70 kVp and VMI with the highest CNR were performed with the two-tailed t test or Kruskal–Wallis test. Results The 40-keV images offered the highest CNR among VMIs. Aortic attenuation at 70 kVp was significantly lower than that at 40 keV (p Conclusion Aortic attenuation at 70 kVp with 180 mg I/kg was lower than that of VMI at 40 keV, and the objective and subjective image qualities were higher at 70 kVp than those at 40 keV.