Focal Epileptiform Discharges Can Mimic Electrode Artifacts When Recorded on the Scalp Near a Skull Defect
2018
Breach rhythm, the hallmark of skull defect, is a familiar finding in the electroencephalogram (EEG). A hole in the skull can also give rise to unfamiliar EEG findings. We present 3 patients with a skull defect whose scalp EEG showed focal epileptiform discharges that resembled F4 electrode artifacts—a 23-year-old man with a right-sided craniectomy for traumatic brain injury, a 63-year-old woman with a history of bifrontal craniectomy and meningioma resection, and a 77-year-old woman who had a right hemicraniectomy for a life-threatening subdural hematoma. In all 3 patients, the F4 electrode was directly above or near a skull defect, and scalp EEG showed phase-reversing waves in FP2-F4 and F4-C4 with no clear-cut “physiological field” (even when the EEG was displayed at a higher sensitivity). In the first 2 patients, the technologist tried to eliminate the “electrode artifacts” by cleaning the scalp thoroughly, replacing the F4 electrode, and maintaining electrode impedance between 2 and 5 kΩ. These measu...