Sensitive helium leak detection in a deuterium atmosphere using a high-resolution quadrupole mass spectrometer

1996 
Abstract In fusion machines, realizing a high-purity plasma is a key to improving the plasma parameters. Thus, leak detection is a necessary part of reducing the leak rate to a tolerable level. However, a conventional helium ( 4 He) leak detector is useless in fusion machines with a deuterium (D 2 ) plasma, because retained D particles on the first walls release D 2 for a long period and the released D 2 interferes with the signals from the leaked 4 He due to the near identical masses of 4.0026 u ( 4 He) and 4.0282 u (D 2 ). A high-resolution quadrupole mass spectrometer (HR-QMS) that we have recently developed, can detect a 4 He + population as small as 10 −4 peak in a D 2 atmosphere. Thus, the HR-QMS has been applied to detect 4 He leaks. To improve the minimum detectable limit of 4 He leak, a differentially pumped HR-QMS analyzer was attached to a chamber of the 4 He leak detector. In conclusion, the improved 4 He leak detector could detect 4 He leaks of the order of 10 −10 Pa · m 3 /s in a D 2 atmosphere.