Coronary CT angiography in patients with implanted cardiac devices: initial experience with the metal artefact reduction technique

2016 
Objective:Single-energy metal artefact reduction (SEMAR), a new technique that can now be used in routine CT examinations, has recently become applicable to volume data acquired with electrocardiography gating. We evaluated the effect of this technique on the visualization of the coronary arteries in patients harboring cardiac devices.Methods:We subjected 8 patients (7 males, 1 female; mean age 65.5 ± 11.3 years) with implanted cardiac devices to coronary CT angiography on a 320-slice CT scanner (Aquilion ONE Vision™; Toshiba Medical Systems Corp., Tokyo, Japan). Image data sets were reconstructed with and without SEMAR. Two radiologists visually evaluated the image quality based on metal artefacts from the electronic device leads using a four-point scale (1 = vessel not visible to 4 = minimal or no metal artefacts). Images with a score of 3 or 4 were considered diagnostic.Results:In both SEMAR and non-SEMAR data sets, 94 coronary artery segments were available for evaluation. Without SEMAR, 11 segments (...