Preanalytical stringency: what factors may confound interpretation of thiopurine S-methyl transferase enzyme activity?

2013 
BackgroundMeasurement of red blood cell thiopurine S-methyl transferase (TPMT) enzyme activity before commencing thiopurine therapy is recommended to avoid severe bone marrow suppression in TPMT-deficient patients. Patient's samples go through preanalytical transportation and storage steps before measurement. We studied patient's TPMT activity sample data to assess the effect of preanalytical variables including transportation time.MethodsA total of 8524 TPMT enzyme activity analyses were conducted from 2002 to 2010 in a single laboratory, with samples sent from seven centres throughout New Zealand. TPMT activity was correlated with time of arrival at the reference laboratory, patient gender and age and centre of sample collection.ResultsThe 6348 (74%) selected TPMT measurements that fulfilled selection criteria ranged from 0 to 25.8 IU/mL. Median delay to sample analysis was 42 h. Median TPMT activity was significantly lower for all centres compared with the reference centre (P < 0.001). Delay in sample ...