Comparison of radiation dose and image quality of triple-rule-out computed tomography angiography between conventional helical scanning and a strategy incorporating sequential scanning.

2011 
Triple-rule-out computed tomographic angiography (TRO CTA), performed to evaluate the coronary arteries, pulmonary arteries, and thoracic aorta, has been associated with high radiation exposure. The use of sequential scanning for coronary computed tomographic angiography reduces the radiation dose. The application of sequential scanning to TRO CTA is much less well defined. We analyzed the radiation dose and image quality from TRO CTA performed at a single outpatient center, comparing the scans from a period during which helical scanning with electrocardiographically controlled tube current modulation was used for all patients (n = 35) and after adoption of a strategy incorporating sequential scanning whenever appropriate (n = 35). Sequential scanning was able to be used for 86% of the cases. The sequential-if-appropriate strategy, compared to the helical-only strategy, was associated with a 61.6% dose decrease (mean dose-length product of 439 mGy × cm vs 1,144 mGy × cm and mean effective dose of 7.5 mSv vs 19.4 mSv, respectively, p