3-T magnetic resonance for determination of aortic valve area: a comparison to echocardiography.

2014 
AbstractObjectives. For evaluation of aortic valve area (AVA), transthoracic echocardiography (TTE) is the method of choice. Cardiac magnetic resonance (CMR) at 1.5-Tesla is an alternative. The aim of the study was to check whether quantification of whole range of AVA without severe aortic stenosis is possible and reliable in higher magnetic field strength, and also including a comparison to TTE. Methods. In 3-T CMR phase contrast sequences were assessed above aortic valve and left ventricular output tract. AVA was calculated using the continuity equation. Planimetric analysis of AVA was performed in magnitude images. TTE was used as reference method for graduation of AVA. Results. Totally 48 patients (64 ± 18 years) without severe aortic valve stenosis were prospectively enrolled. In CMR planimetric AVA was 2.5 ± 1.3 cm2 and calculated AVA 2.4 ± 1.3 cm2, whereas AVA in TTE was 1.9 ± 1.1 cm2. Planimetric and calculated AVA in CMR and also AVA in CMR and TTE showed good correlation (r = 0.97, 0.92, respect...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    28
    References
    4
    Citations
    NaN
    KQI
    []