Biosensor-based fragment screening using FastStep injections

2010 
Abstract We have developed a novel analyte injection method for the SensiQ Pioneer surface plasmon resonance-based biosensor referred to as “FastStep.” By merging buffer and sample streams immediately prior to the reaction flow cells, the instrument is capable of automatically generating a two- or threefold dilution series (of seven or five concentrations, respectively) from a single analyte sample. Using sucrose injections, we demonstrate that the production of each concentration within the step gradient is highly reproducible. For kinetic studies, we developed analysis software that utilizes the sucrose responses to automatically define the concentration of analyte at any point during the association phase. To validate this new approach, we compared the results of standard and FastStep injections for ADP binding to a target kinase and a panel of compounds binding to carbonic anhydrase II. Finally, we illustrate how FastStep can be used in a primary screening mode to obtain a full concentration series of each compound in a fragment library.