Adaptation and validation of a commercial head phantom for cranial radiosurgery dosimetry end-to-end audit.

2017 
Objective:To adapt and validate an anthropomorphic head phantom for use in a cranial radiosurgery audit.Methods:Two bespoke inserts were produced for the phantom: one for providing the target and organ at risk for delineation and the other for performing dose measurements. The inserts were tested to assess their positional accuracy. A basic treatment plan dose verification with an ionization chamber was performed to establish a baseline accuracy for the phantom and beam model. The phantom and inserts were then used to perform dose verification measurements of a radiosurgery plan. The dose was measured with alanine pellets, EBT extended dose film and a plastic scintillation detector (PSD).Results:Both inserts showed reproducible positioning (±0.5 mm) and good positional agreement between them (±0.6 mm). The basic treatment plan measurements showed agreement to the treatment planning system (TPS) within 0.5%. Repeated film measurements showed consistent gamma passing rates with good agreement to the TPS. Fo...