Evaluating elevated release liner adhesion of a transdermal drug delivery system (TDDS): A study of Daytrana™ methylphenidate transdermal system

2011 
Background: Complaints from healthcare providers that the adhesive on the Daytrana™ methylphenidate transdermal drug delivery system (TDDS) adhered to the release liner to such an extent that the release liner could not be removed prompted this study. Daytrana™ has a packaging system consisting of a moisture-permeable pouch contained within a sealed tray containing a desiccant; the tray is impermeable to ambient moisture. The objective of this project was to determine if the Daytrana™ packaging system influenced the difficulty in removing the release liner.Method: Both a sealed tray and an open tray containing sealed pouches were placed into an environmental chamber at 25°C and 60% relative humidity for 30 days; afterwards, release liner removal testing using a peel angle of 90° and a peel speed of 300 mm/min was performed.Results: TDDS from open chamber trays required less force to remove the release liner than did TDDS from closed chamber trays. For the 10 mg/9 h TDDS and the 15 mg/9 h TDDS (the dosages...