Undetected oesophageal perforation and feeding-tube misplacement

2014 
This is a case report of an electromagnetically (EM)-guided Cortrak feeding tube that perforated the lower oesophagus and was not detected by the EM trace or by plain X-ray. Misplacement was diagnosed from computed tomography (CT) following injection of radio-contrast medium down the tube. Recommendations are offered for use of the EM trace in patients at high risk of oesophageal perforation.