Ultrasensitive Detection of Cancer Biomarkers in the Clinic by Use of a Nanostructured Microfluidic Array

2012 
Multiplexed biomarker protein detection holds unrealized promise for clinical cancer diagnostics due to lack of suitable measurement devices and lack of rigorously validated protein panels. Here we report an ultrasensitive electrochemical microfluidic array optimized to measure a four-protein panel of biomarker proteins, and we validate the protein panel for accurate oral cancer diagnostics. Unprecedented ultralow detection into the 5–50 fg·mL–1 range was achieved for simultaneous measurement of proteins interleukin 6 (IL-6), IL-8, vascular endothelial growth factor (VEGF), and VEGF-C in diluted serum. The immunoarray achieves high sensitivity in 50 min assays by using off-line protein capture by magnetic beads carrying 400 000 enzyme labels and ∼100 000 antibodies. After capture of the proteins and washing to inhibit nonspecific binding, the beads are magnetically separated and injected into the array for selective capture by antibodies on eight nanostructured sensors. Good correlations with enzyme-linke...